The tester is used for testing the content of Figure 16 ~ 27 in the latest edition of GB9706.1 (idt IEC 60601-1) "Medical electrical equipmentPart 1: General requirements for safety". Operators can finish different combination of switches and select the test in normal condition or single fault condition. Customized converter box is available for high-power or three-phase equipment.
Output voltage: adjustable from 0V to 250V (AC); with LED digital display; error: within ±2% of reading
Output power: 500W and 1kW
Input impedance: compliant with the latest edition of GB 9706.1
Measuring range: 0.5uA~10mA; DC -1MHz True RMS; with LED digital display; error: within ±5% of reading
Leakage current alarm threshold: 0.01 / 0.05 / 0.1 / 0.5 / 1 / 2.5 / 5 / 10 (mA) (8 options)
External voltage: adjustable from 0V to 250V (AC); with LED digital display
Protective impedance: TYPE B: 10K; TYPE BF (120V): 22K; TYPE BF (240V): 47K; TYPE CF: 100K