Features:
Parameters:
|
Power Supply |
100 – 260 VAC, 50HZ, 120W |
|
Operation Pressure |
100-10000 mbar (1.5-145psi) ; |
|
Unit |
mbar |
|
Operation Condition |
Ambient Temperature:+5℃ ~ +40℃;Relative Humidity:10-80% |
|
Dimension(mm) |
400(Depth)*240(Width)*270(Height1)*220(Height2) |
|
Function |
Manual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test; Pressure Holding Test;Diffusion FLow Test;Water Intrusion Test;Ultrafitration Membrane Test |
|
Test Accuracy |
Upstream Volume Test:≤± 4%;Bubble Point Test:≤± 50mbar ;Diffusion Flow Test:≤± 4% ;Water Immersion:≤± 0.01ml |
|
Test Scope |
BP:100-8000mbar DF:1-600ml/min WI:0.01-100ml/min |
|
Applied Range |
Symmetric and asymmetric membrane test, needle filter, filter bag, flat filter, filter cartridge (big filter volume ), ultrafiltration membrane, ultrafiltration column, a variety of irregular filters |
|
Authority Management |
4 level User Name and Password to Login,electronic signature |
|
Reservation Solution |
120 sets |
|
Audit trail |
customization |
|
Print Function |
Built-in printer; |
|
Historical Records |
5000 sets unlimited sets records |
|
Records Copy |
Support USB disk to export (Include test curve) |
|
Display Screen |
7 inch, color touch screen |
|
Serial Connection |
RS232/USB |
|
Language |
Chinese/English |
|
Work Mode |
Online/Offline |
|
Applicable Environment |
Above D level |
|
Weight |
8KG |
|
Signal output |
(4-20)MA,RS485,12V Alarm output |